CVE-2017-11579

Description

In the most recent firmware for Blipcare, the device provides an open Wireless network called "Blip" for communicating with the device. The user connects to this open Wireless network and uses the web management interface of the device to provide the user's Wi-Fi credentials so that the device can connect to it and have Internet access. This device acts as a Wireless Blood pressure monitor and is used to measure blood pressure levels of a person. This allows an attacker who is in vicinity of Wireless signal generated by the Blipcare device to easily sniff the credentials. Also, an attacker can connect to the open wireless network "Blip" exposed by the device and modify the HTTP response presented to the user by the device to execute other attacks such as convincing the user to download and execute a malicious binary that would infect a user's computer or mobile device with malware.

CvssV3 impact

BaseSeverity

HIGH

ConfidentialityImpact

HIGH

AttackComplexity

LOW

Scope

UNCHANGED

AttackVector

ADJACENT_NETWORK

AvailabilityImpact

NONE

IntegrityImpact

LOW

PrivilegesRequired

NONE

BaseScore

7.1

VectorString

CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:L/A:N

Version

3.0

UserInteraction

NONE

CvssV2 impact

Version

2.0

VectorString

AV:A/AC:L/Au:N/C:P/I:P/A:N

AccessVector

ADJACENT_NETWORK

AccessComplexity

LOW

Authentication

NONE

ConfidentialityImpact

PARTIAL

IntegrityImpact

PARTIAL

AvailabilityImpact

NONE

BaseScore

4.800000190734863